Evaluation of the static and dynamic assistive torque of a passive upper limb occupational exoskeleton

Published online by Cambridge University Press:  15 April 2025

Etienne Ricard*
Affiliation:
Département Sciences Appliquées au Travail et aux Organisations, Institut National de Recherche et Sécurité (INRS), Vandoeuvre-les-Nancy, France ComBO research team, Université de Rennes, Inria, CNRS, IRISA-UMR 6074, Rennes, France
Chris Hayot
Affiliation:
Département Sciences Appliquées au Travail et aux Organisations, Institut National de Recherche et Sécurité (INRS), Vandoeuvre-les-Nancy, France
Isabelle Clerc-Urmès
Affiliation:
Département Sciences Appliquées au Travail et aux Organisations, Institut National de Recherche et Sécurité (INRS), Vandoeuvre-les-Nancy, France
Laurent Claudon
Affiliation:
Département Sciences Appliquées au Travail et aux Organisations, Institut National de Recherche et Sécurité (INRS), Vandoeuvre-les-Nancy, France
Kévin Desbrosses
Affiliation:
Département Sciences Appliquées au Travail et aux Organisations, Institut National de Recherche et Sécurité (INRS), Vandoeuvre-les-Nancy, France
Charles Pontonnier
Affiliation:
ComBO research team, Université de Rennes, Inria, CNRS, IRISA-UMR 6074, Rennes, France
*
Corresponding author: Etienne Ricard; Email: [email protected]

Abstract

Adjusting the assistive torque of upper limb occupational exoskeletons is essential to optimize their effectiveness and user acceptance in companies. This adjustment enables a balance to be struck between the expected benefits and potential undesirable effects associated with their use, particularly for the shoulder joint, which is sensitive to the balance of forces. Despite this, no study has yet evaluated these assistive torques in static and dynamic conditions representative of work situations. The aim of this article is therefore to evaluate these assistive torques under these two conditions, using an isokinetic dynamometer. Angular velocities ranging from 0 to 240°/s and four levels of assistance were investigated. The results showed that the maximum assistive torques in flexion (energy restitution phase) were lower than those in extension (tensioning phase) by 20 to 36% and were median in static conditions. It was also observed that the level of assistance and the exoskeleton opening angles had a strong impact on the assistive torques, unlike the angular velocity in dynamic conditions, which had a minimal effect. Quantifying these assistive torques is crucial for assessing their biomechanical impact and adjusting the exoskeleton’s assistance to the operator and the task performed.

Type
Research Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press

1. Introduction

Passive upper limb exoskeletons (ULE) assisting the shoulder are mainly used for manual materials handling (MMH) and overhead work (OHW) tasks involving dynamic lifting or static holding of arms at height. In such situations, these technologies are designed to reduce muscular stress of the shoulders, which are the site of many musculoskeletal disorders (Barthelme et al., Reference Barthelme, Sauter, Mueller and Liebers2021). The operating principle of these exoskeletons is based on an external torque assisting the muscles involved in arm elevation. Many studies have evaluated the biomechanical effects of wearing ULE through experimental and modeling approaches (Moeller et al., Reference Moeller, Krell-Roesch, Woll and Stein2022; Ma et al., Reference Ma, Zhang, Choi and Xiong2023). These studies generally showed that the use of ULE significantly reduces the activity of the shoulder agonist muscles during arm elevation (Theurel and Desbrosses, Reference Theurel and Desbrosses2019; Zhou and Zheng, Reference Zhou and Zheng2021). However, there is considerable heterogeneity in the experimental results supposing a complex interaction between the exoskeleton, the user, and the task performed. Moreover, some studies reported the presence of undesirable effects: overactivations for shoulder antagonistic muscles in arm elevation, as well as for postural or rotator cuff muscles (Moeller et al., Reference Moeller, Krell-Roesch, Woll and Stein2022; Kranenborg et al., Reference Kranenborg, Greve, Reneman and Roossien2023), excessive pressure at the human–exoskeleton interface (Linnenberg and Weidner, Reference Linnenberg and Weidner2022), perceived discomfort, and limited ease of use (Kranenborg et al., Reference Kranenborg, Greve, Reneman and Roossien2023). Among the many hypotheses that could explain these observations (variations in effects and undesirable effects), the question of optimal assistance must be asked, both statically and dynamically. This is all the more true as the levels of assistance are rarely specified in previous studies, which does not allow for the impact of setting to be taken into account.

Conversely, some researchers have focused on the effect of adjusting the exoskeleton’s assistance in order to determine the best assistance setting. These studies have shown that increasing the level of assistance leads to a decrease in agonistic muscle activity and an increase in antagonistic muscle activity during shoulder elevation (Van Engelhoven et al., Reference Van Engelhoven, Poon, Kazerooni, Barr, Rempel and Harris-Adamson2018; de Vries et al., Reference de Vries, Murphy, Könemann, Kingma and de Looze2019; Sänger et al., Reference Sänger, Yao, Schubert, Wolf, Molz, Miehling, Wartzack, Gwosch, Matthiesen and Weidner2022). The adjustment of the assistance prevented an increase in shoulder reaction force (Sylla et al., Reference Sylla, Bonnet, Colledani and Fraisse2014) or rotator cuff muscle activity (Seiferheld et al., Reference Seiferheld, Frost, Krog, Skals and Andersen2022) and reduced the perception of effort (Van Engelhoven et al., Reference Van Engelhoven, Poon, Kazerooni, Barr, Rempel and Harris-Adamson2018). The preferred settings varied depending on the weight handled and on the user perception (Sänger et al., Reference Sänger, Yao, Schubert, Wolf, Molz, Miehling, Wartzack, Gwosch, Matthiesen and Weidner2022), underlining the importance of precise settings for each task and each user. Thus, the assistance level appears to be an important parameter that determines the exoskeleton effects, the mechanisms of which can only be studied if the quantification of assistance levels can be evaluated and modeled to match the requirements of the task. However, little is known about the mechanical behavior and recommended settings of each ULE in relation to specific task/morphology. Therefore, evaluating the assistive torques, and their variations during the movement (i.e., for different angles or angular velocities), seems to be a fundamental requirement to optimize the use of the exoskeleton. This knowledge would make it possible to adapt the assistive torque to the operator and the task carried out so that the exoskeleton is the most effective in terms of benefits, acceptance, and reducing undesirable effects.

Several studies have already proposed experimental approaches for evaluating the assistive torque of ULE and passive back exoskeletons. Methodologically, most of them have used load cells or force sensors (Koopman et al., Reference Koopman, Kingma, Faber, de Looze and van Dieën2019; Asgari et al., Reference Asgari, Phillips, Dalton, Rudl and Crouch2020; Hartmann et al., Reference Hartmann, Rinaldi, Taira and Forner-Cordero2021; Van Harmelen and Schnieders, Reference Van Harmelen and Schnieders2022; McCann et al., Reference McCann, Hohimer, O’Neill, Young, Bertoldi and Walsh2023). These studies provided information on the torques transmitted by exoskeletons to operators. Some of them reported higher torques during the tensioning phase of the exoskeleton assistance than during the restitution phase (i.e., hysteresis behavior). However, these studies have certain limitations. They are limited to low angular speeds, less than 20°/s, linked to customized test benches sensitive to vibration and with low acquisition frequencies. The position of the sensor depends on the experimenter and the design of the exoskeleton, making it difficult to reproduce these protocols. In addition, the reliability of the measurements has not been assessed. To overcome these limitations, other authors have characterized the exoskeleton torque–angle relationship using an isokinetic dynamometer. This device was used to characterize the assistive torques of two back exoskeletons in static and dynamic conditions, also highlighting hysteresis behavior (Madinei et al., Reference Madinei, Kim, Park, Srinivasan and Nussbaum2022). However, for ULE, only static conditions were evaluated and modeled using polynomial equations (Watterworth et al., Reference Watterworth, Dharmaputra, Porto, Cort and La Delfa2023), though most MMH situations involve dynamic movements.

Thus, this study aims to evaluate and model the assistive torque of an ULE using an isokinetic dynamometer in static and dynamic conditions. In particular, the study seeks to answer two questions (Q):

Q1: What assistive torque does the exoskeleton provide as a function of its settings (assistance level) and conditions of use (exoskeleton arm elevation angle, angular velocity and rotation direction of the movement)?

Q2: How to model these assistive torque profiles using equations, to integrate them in the design phase of a workstation? Knowing and modeling exoskeleton assistance as a function of setting and joint configuration has two major benefits. The first is to be able to propose optimal conditions of use for a given set of tasks. The second is to use the modeling in motion analysis tools to understand and quantify the biomechanical impact of the system on operator task performance.

2. Material and methods

2.1. Upper limb exoskeleton

The Ottobock shoulder (Ottobock SE & Co. KGaA, Duderstadt, Germany) was assessed in this study (Figure 1). It is a passive ULE designed to assist the user during OHW tasks. It weighs 2 kg and assists elevation of each operator arm with a system composed of a bungee cord (passive actuator) and rigid segments (exoskeleton arm and adjustable support bar). The passive actuator is put under tension with the weight of the user’s arm or when lowering the exoskeleton arm. This system generates a torque depending on the elevation angle of the operator arm (Finke et al., Reference Finke, Kroll-Orywahl, Tüttemann, Bremer, Merker, Mizera, Mladek, Röschel, Schirrmeister, Vogel, Bornmann, Lanzinger, Käsmann and Schwedhelm2022). The angular range of motion of the exoskeleton arm is 155°. Assistance level is adjustable by an endless screw and four levels of assistance are indicated as shown in Figure 1: a minimum (Min), two intermediate (Int1 and Int2), and a maximum level (Max). These four levels of assistance were tested in this study. To follow the kinematics of the shoulder, an adjustable support bar is connected to a hip belt with a ball joint. The adjustable support bar was mounted parallel to the passive actuator and is connected to the exoskeleton arm by a hinge joint. Finally, a cuff links the operator’s arm to the exoskeleton arm. The structure can be adapted for operators between 1.60 m and 1.90 m tall.

Figure 1. Ottobock shoulder exoskeleton with its different elements and assistance levels to the study: minimum in light blue (Min), two intermediate levels in green and magenta (Int 1 and Int 2), and a maximum in dark blue (Max) (Exoskeleton Report, 2022).

2.2. Isokinetic dynamometer

The exoskeleton torque–angle relationship was quantified using the isokinetic dynamometer HUMAC NORM (Computer Sports Medicine, Inc., Stoughton, MA, USA) in isometric mode (for static conditions) and in continuous passive motion mode (for dynamic conditions) at different angular velocities. In this latter, the angular velocities were regulated by the integrated servo-motor of the isokinetic dynamometer. HUMAC NORM was validated to evaluate torque for both conditions in many studies (de Araujo Ribeiro Alvares et al., Reference de Araujo Ribeiro Alvares, Rodrigues, de Azevedo Franke, da Silva, Pinto, Vaz and Baroni2015; Habets et al., Reference Habets, Staal, Tijssen and van Cingel2018; Estrázulas et al., Reference Estrázulas, Estrázulas, de Jesus, de Jesus, da Silva and Libardoni dos Santos2020). The device settings were defined using the HUMAC 2015 software. The continuous passive motion mode was set with a cushion value of 0 corresponding to the slowest deceleration (softest stop) to avoid hard stop at the amplitude limits. The isokinetic dynamometer was attached to the exoskeleton’s cuff equipped with a stabilizer roller pad, as if it was a human arm. In addition, the rotation axes of the isokinetic dynamometer and the exoskeleton were visually aligned by means of a laser. The exoskeleton was securely fastened to the dynamometer seats using straps to prevent any unwanted movement (Figure 2). The torque, angle, and angular velocity data were recorded via the analog output of the HUMAC NORM and sampled at 1200 Hz.

Figure 2. Detailed view of the test bench with the Ottobock shoulder exoskeleton positioned in its maximum opening angle (155°) on the HUMAC NORM isokinetic dynamometer.

2.3. Experimental protocol

Only one experimenter performed the measurements. In static condition (STA), the exoskeleton assistive torque was recorded over a period of 5 sec for a first measurement at 0° and for every 10° from 5 to 155° (i.e., 17 measuring angles). Five repetitions were performed for each angle and each of the four exoskeleton assistance levels. In dynamic conditions, the torque was continuously recorded during the movement of the exoskeleton arm over its full range of motion in both directions: from 0 to 155°, representing a shoulder flexion movement (FLE) if the exoskeleton was worn by a person, and from 155 to 0°, representing a shoulder extension movement (EXT). The positioning of the exoskeleton on the isokinetic dynamometer, as well as the corresponding positioning on an operator, are described in Supplementary Material A. Angular velocities ranging from 20 to 240°/s with an incremental step of 20°/s (i.e., 12 angular velocities) were completed. Five repetitions of flexion/extension movements were successively performed for each angular velocity and each of the four exoskeleton assistance levels. A pause of 2 sec was set between flexion and extension motion. Additionally, to isolate the exoskeleton torque from the net weight of the dynamometer arm, repetitions were carried out without the exoskeleton for each condition. During this protocol, angles, angular velocities, and levels of assistance were randomized.

2.4. Data processing

In order to determine the assistive torques provided by the exoskeleton, the raw data were low-pass filtered (fourth order bidirectional Butterworth filter, cut-off frequency = 3 Hz). The cut-off frequency was determined using the residual analysis method (Winter, Reference Winter2009). For STA torques, each trial was averaged over 5 sec. For FLE and EXT torques, each trial was averaged every degree (from 0 to 1° and 154 to 155°). Therefore, the torque value corresponding to 0° is the average of the torque recorded within the interval [0°; 1°] and so on. For this reason, torque profiles were calculated for the static and dynamic conditions from 0 to 154°. The repeatability of the trials without exoskeleton under static and dynamic conditions, respectively, was checked using Bland–Altman diagrams (Martin Bland and Altman, Reference Martin Bland and Altman1986) to average these trials of each condition. These averages were subtracted from the total torque measured in the corresponding condition with exoskeleton. To compare the maximum torques and their positions in STA with FLE and EXT, cubic spline interpolation was performed on the 17 angles of isometric measurements. The data processing was carried out using Matlab (version R2023b).

2.5. Statistical analysis

The results were presented as the mean $ \pm $ standard deviation for the five STA trials, with FLE and EXT for the four levels of assistance. FLE and EXT were presented for the 12 angular velocities. To compare each level of assistance, angular velocity and STA, FLE, and EXT conditions, a T test or a Wilcoxon rank sum test was used on the maximum torque and their positions, as appropriate. To determine assistive torque equations as a function of the considered opening angle, the level of assistance and angular velocity, according to the profiles considered for each of the STA, FLE, and EXT conditions, an ANOVA model was first performed. Main effects and their interactions were tested. Significant effects were then analyzed using Bonferroni post hoc comparison tests. The effect size was evaluated by means of Cohen’s d values for significant post hoc comparisons. The effect size was considered small ( $ 0.2\le d<0.5 $ ), medium ( $ 0.5\le d<0.8 $ ), or large ( $ 0.8\le d $ ) according to Cohen (Reference Cohen1992). Only results with very small effect size ( $ d<0.2 $ ) were not considered.

Then, piecewise polynomial regression models were used to take into account the STA, FLE, and EXT profiles and conditions (Ostertagová, Reference Ostertagová2012; Watterworth et al., Reference Watterworth, Dharmaputra, Porto, Cort and La Delfa2023). Each regression model was evaluated at polynomial orders from 1 to 3. Following preliminary work to improve the adjusted R 2 ( $ {R}_{adj}^2 $ ) by 1% per added parameter, a maximum of six parameters were fixed. These models were selected with the highest $ {R}_{adj}^2 $ and the lowest Akaike information criterion (AIC) (Akaike, Reference Akaike1974; Sugiura, Reference Sugiura1978; Hurvich et al., Reference Hurvich, Simonoff and Tsai1998; Wagenmakers and Farrell, Reference Wagenmakers and Farrell2004). Additionally, the minimal negative values of assistive torque obtained were utilized as indicators for model selection (Looft and Law, Reference Looft and Law2015).

Alpha risk was fixed to 5% for all analyses. The statistical analysis was carried out using STATA software (Stata 18, StataCorp., College Station, TX, USA).

3. Results

3.1. Exoskeleton assistive torque profiles

A descriptive presentation of the exoskeleton torque profile according to the exoskeleton opening angles for the four levels of assistance is shown in Figure 3. Bell-shaped curves can be observed. In STA, two profiles are identified on angular ranges [0°; 25°] and ]25°; 154°]. In FLE (all angular velocities combined), two profiles are identified on angular ranges [0°; 7°] and ]7°; 154°] and in EXT (all angular velocities combined), one profile is identified on the full angular range [0°; 154°].

Figure 3. Torque–angle relationship of the Ottobock shoulder under different conditions: STA (left), FLE (center), and EXT (right) for the four assistance levels (Min, Int1, Int2, Max), represented by the mean with the standard deviation in the band. The dynamic conditions FLE and EXT are represented for the 12 angular velocities from 20 to 240°/s.

3.2. Effect of assistance levels, velocity, and rotation direction

According to the previously defined statistical criteria, no effect of angular velocity in FLE on maximum torques and their positions was considered ( $ d<0.20 $ ). However, in EXT, the maximum torque decreased with faster angular velocities. Small effects size were obtained, from 80 to 240°/s compared to 20°/s ( $ 0.20\le d\le 0.32 $ ), from 120 to 140°/s compared to 40°/s ( $ 0.20\le d\le 0.23 $ ), and for 120°/s compared to 60°/s (d = 0.20). For angular positions, the positions of the maximum torque at 240°/s were greater ( $ 74.30{}^{\circ}\pm 3.70{}^{\circ} $ ) than those obtained for 100 ( $ 68.15{}^{\circ}\pm 6.42{}^{\circ} $ ) and 120°/s ( $ 68.00{}^{\circ}\pm 6.26{}^{\circ} $ ) with large effects size ( $ d>0.8 $ ).

Raising the assistance levels resulted in an increase in the maximum torques for STA, FLE, and EXT conditions, with large effect size ( $ d>0.8 $ ) (Figure 4). However, it did not affect the angular positions for STA and EXT. But in FLE, there was a decrease in angular positions of maximum torque with a small effect size between Min and Max (d = 0.48), Int1 and Int2 (d = 0.49), and a moderate effect between Min and Int2 (d = 0.51) (Figure 4).

Figure 4. Boxplot representation to compare the exoskeleton’s maximum torques (top) and their angular positions (bottom) for the four assistance levels (Min, Int1, Int2, and Max) in STA, FLE, and EXT including 5 trials and 12 angular velocities.

Concerning the maximum torque values, greater values were observed in EXT than in STA ( $ 2.41\le d\le 5.22 $ ), and in STA than in FLE ( $ 1.67\le d\le 3.43 $ ) for each level of assistance and angular velocity, as depicted in Figure 4.

Concerning the position of these maximum torques, greater angles were obtained in STA than in FLE ( $ 1.24\le d\le 4.21 $ ), except for the Max level for angular velocities from 40 to 100°/s, and in FLE than in EXT ( $ 0.25\le d\le 1.18 $ ) for a part of level of assistance and angular velocities, as shown in Figure 4.

3.3. Modeling of exoskeleton assistive torque

For the complete profiles in EXT, angular velocities had no significant effect. In FLE, there was a significant effect with a very small effect size where d < 0.2, so the angular velocity was not taken into account for modeling. For STA, FLE, and EXT, only the level of assistance had a large effect size when there was a difference of at least two levels of assistance ( $ d>0.8 $ ). To model the assistive torques, two independent variables were used, where “x” corresponds to the exoskeleton opening angles from 0 to 154° and “y” to the levels of assistance between 1 and 4, where 1 is the Min level, 2 and 3 two intermediate levels, and 4 the Max level. To model STA, two equations were used: STA 1 on an angular range [0°; 25°] and STA 2 on ]25°; 154°]. In FLE, two equations were used: FLE 1 on [0°; 7°] and FLE 2 on ]7°; 154°]. In EXT, one equation was used, EXT 1 on [0°; 154°]. These equations are shown in Table 1 with good fitting indicators, root mean squared error (RMSE) between [0.16 Nm; 0.43 Nm] and $ {R}_{adj}^2 $ [0.882; 0.982]. The representation of these modeling results are provided in Supplementary Material B.

Table 1. Ottobock shoulder assistive torques equations fitted by polynomial regression models for STA, FLE, and EXT conditions. The equations were z = f(x,y), where x corresponds to the considered angle and y the assistance levels between 1 and 4, where 1 is the Min level, 2 and 3 Int1 and Int2 levels, and 4 the Max level, with their fitting indicators, RMSE and $ {\mathrm{R}}_{\mathrm{adj}}^2 $

4. Discussion

4.1. Assessment of exoskeleton assistive torque

The first objective of this study was to evaluate the assistive torque provided by an ULE under static and dynamic conditions.

In static conditions, the torque-opening angle relationships formed bell-shaped profiles. Maximum torque values ranged from 3.3 to 8.2 Nm, depending on the level of assistance, for opening angles between 79 and 94°. These results were similar to the assistive torques provided by the first version of the Ottobock shoulder (i.e., Paexo shoulder) with two modes, single spring and double spring. The maximum assistive torques varied from 3 to 6 Nm for the single spring mode, and from 6 to 11 Nm for the double spring mode. These torques were obtained for opening angles between 90 and 100°, using an isokinetic dynamometer under static conditions (Watterworth et al., Reference Watterworth, Dharmaputra, Porto, Cort and La Delfa2023). The results of our study are also similar to other ULE for which the maximum torques varied between 4 and 13 Nm (Watterworth et al., Reference Watterworth, Dharmaputra, Porto, Cort and La Delfa2023).

In dynamic conditions, the torque-opening angle relationships also formed bell-shaped profiles. Their maximum torques ranged between 2.7 and 8.9 Nm (according to the level of assistance) at corresponding opening angles between 60 and 85°. To our knowledge, there are no other studies in the literature reporting maximum dynamic assistive torques for ULE.

The angular velocities investigated in this study had minimal effect on the assistive torques. Similarly, previous research by Madinei et al. (Reference Madinei, Kim, Park, Srinivasan and Nussbaum2022) observed no effect on provided torque for two back exoskeletons across angular velocities ranging from 20 to 100°/s. This suggests that exoskeleton assistive torques are not significantly influenced by angular velocity, unlike the torque-producing capacity of the shoulder. In fact, the concentric torque production capacity at the shoulder decreased with increasing angular velocity (Otis et al., Reference Otis, Warren, Backus, Santner and Mabrey1990; Whitcomb et al., Reference Whitcomb, Kelley and Leiper1995). The reported reductions were attributed to the muscle force–velocity relationship. Thus, the exoskeleton’s assistive torque could be more beneficial at higher velocity.

The rotation direction of the movement was also investigated in this study. The assistive torques were larger during the extension movement than the flexion movement and median in static. These results, comparing static and dynamic conditions, were also observed for two passive back exoskeletons (Madinei et al., Reference Madinei, Kim, Park, Srinivasan and Nussbaum2022). These differences between static and dynamic conditions are mainly due to the friction forces of the device. These forces were subtracted from the static torques during the restitution phase (FLE) and added to them during the tensioning phase (EXT). In addition, the mechanical characteristics of the elastic band or spring lead to a loss of energy during the restitution phase. This difference in dynamic torque is known as hysteresis. Many studies refer to it for ULE (Asgari et al., Reference Asgari, Phillips, Dalton, Rudl and Crouch2020; McCann et al., Reference McCann, Hohimer, O’Neill, Young, Bertoldi and Walsh2023) or back exoskeletons (Koopman et al., Reference Koopman, Kingma, Faber, de Looze and van Dieën2019; Madinei et al., Reference Madinei, Kim, Park, Srinivasan and Nussbaum2022). Maximum percentages of energy loss have been referenced from 28 to 51% for two back exoskeletons (Van Harmelen and Schnieders, Reference Van Harmelen and Schnieders2022) and 28% for an ULE (Rossini et al., Reference Rossini, De Bock, van der Have, Flynn, Rodriguez-Cianca, De Pauw, Lefeber, Geeroms and Rodriguez-Guerrero2021). The results of this study indicated maximum energy losses of 36, 29, 23, and 20% for Min, Int1, Int2, and Max, respectively. The difference between FLE and EXT is constant at around 1.6 Nm. Compared with the torque production capacity of the shoulder, these results point in the same direction. For the same shoulder elevation angle, the torques generated are higher for an extension movement than for a flexion movement and median in static condition (Koski and McGill, Reference Koski and McGill1994; Harbo et al., Reference Harbo, Brincks and Andersen2012).

Inverse bell-shaped profiles with low assistive torques were obtained on STA [0°; 25°] and FLE [0°; 7°] conditions. These profiles were also observed for the first version of the Ottobock shoulder (i.e., Paexo shoulder) and for other ULE (Watterworth et al., Reference Watterworth, Dharmaputra, Porto, Cort and La Delfa2023). They result from the maximum tension of the force transmission cable when the opening angle is low. Indeed, in this configuration, the tension of the elastic does not allow the system to passively maintain itself at 0°. Therefore, there is a torque greater than 0 Nm at this angle. Then, the temporary reduction in torque for the following angles is probably due to the configuration of the pulleys designed to assist at the desired angle (Rossini et al., Reference Rossini, De Bock, van der Have, Flynn, Rodriguez-Cianca, De Pauw, Lefeber, Geeroms and Rodriguez-Guerrero2021; Kim et al., Reference Kim, Kim, Jung, Lee and Bae2022). According to Kim et al., Reference Kim, Kim, Jung, Lee and Bae2022, they enable permanent contact to be maintained between the exoskeleton cuff and the user’s arm.

4.2. Modeling of exoskeleton assistive torque

The second objective of this study was to model assistive profiles obtained during the evaluation of the assistive torques. The assistive profiles were modeled as a function of two selected input parameters, the opening angle of the exoskeleton “x” and the setting of its level of assistance “y.” In static conditions, piecewise polynomial modeling was performed to fit the two assistance profiles. These equations are STA 1 for angles ranging from [0°; 25°] and STA 2 for angles ranging from ]25°; 154°]. These models have high $ {R}_{adj}^2 $ values of 0.88 and 0.96, respectively, attesting to the accurate modeling of the assistance profiles. To our knowledge, Watterworth et al. (Reference Watterworth, Dharmaputra, Porto, Cort and La Delfa2023) is the only study to have modeled the assistive torque of ULE as a function of the shoulder angle elevation in static condition. An equation per level of assistance was performed with similar $ {R}_{adj}^2 $ . They hypothesized that the shoulder elevation angle was close enough to the exoskeleton opening angle to be included in the equation. Conversely, in this study, if the opening angle of the exoskeleton is not known, it is possible to use the shoulder elevation angle as an input by making the same assumption. These equations will enable us to obtain the output assistive torques for a given level of assistance and angle in static conditions.

In dynamic conditions, a piecewise polynomial modeling was performed for FLE to fit the two assistance profiles, FLE1 and FLE2 on the angular ranges [0°; 7°] and ]7°; 154°], respectively. One equation was performed for EXT, EXT1 over the entire profile. These equations integrate the mechanical properties of the ULE under dynamic conditions for angular velocities of 20 to 240°/s corresponding to the majority of MMH tasks and part of the OHW tasks. Despite taking into account all the angular velocities, the fitting is very good with $ {R}_{adj}^2 $ between [0.93; 0.98]. This study is the first to model exoskeleton assistive profiles under dynamic conditions. These five equations obtained in static and dynamic conditions can be used to obtain the assistive torques generated by the ULE during a work situation.

In the future, this work could help to adjust exoskeleton assistance according to the operator and the task. Based on these equations, it would be possible to develop a computer tool to adjust the exoskeleton’s assistance according to the shoulder torque for different work situations and anthropometry of operators (de Leva, Reference de Leva1996). Another approach for researchers would consist of estimating the effects of the ULE via digital human modeling (van der Have et al., Reference van der Have, Rossini, Rodriguez-Guerrero, Van Rossom and Jonkers2022; Delgado-Llamas et al., Reference Delgado-Llamas, Marín-Boné and Marín-Zurdo2023). Various assistive torques could be applied to the human biomechanical model at the shoulder. Thus, using a calculation procedure (inverse dynamics for example), it would be possible to calculate the forces at the shoulder joint and optimize the level of assistance to minimize joint stresses. In addition, a constrained optimization procedure could be used to calculate muscle forces and joint contact forces. These estimated musculoskeletal loads should enable an optimal balance to be found between the beneficial and undesirable effects of wearing the exoskeleton (Seiferheld et al., Reference Seiferheld, Frost, Krog, Skals and Andersen2022). Current modeling approaches do not take into account the mechanical behavior of exoskeletons, such as hysteresis (Ma et al., Reference Ma, Zhang, Choi and Xiong2023; Scherb et al., Reference Scherb, Wartzack and Miehling2023). The protocol proposed in this study could be a way of improving the accuracy of digital human modeling and thus enabling the simulation of the exoskeleton effect during the design phase of a workstation.

Some limitations of this study should also be mentioned. The ULE was characterized under optimal conditions of use, with perfect alignment with the rotation axes of the isokinetic dynamometer. Real operating usage conditions do not guarantee perfect assistance efficiency, and the full assistive torque is not transmitted to the operator. The conditions of use may entail misalignments due to the system’s freedom of movement, slippages at the interface, contact forces, and so forth. In addition, the transfer of force from the assist torque to the pelvis was not measured and therefore could not be modeled. Furthermore, the protocol of this study included only one model of this ULE. However, it could be possible that differences in torque generation could exist between the same ULE models when they are manufactured. The reproducibility of this protocol for other exoskeletons could be limited by their design, particularly for exoskeletons with many degrees of freedom.

4.3. Practical application

ULE were designed to compensate for shoulder elevation torques through a maximum assistive torque at approximately 90°. Indeed, raising an outstretched arm with a 2.25 kg tool also generates a bell-shaped torque with a maximum value between 16 and 24 Nm (depending on the operator’s morphology) at 90° (Van Engelhoven et al., Reference Van Engelhoven, Poon, Kazerooni, Rempel, Barr and Harris-Adamson2019). Furthermore, the shoulder’s maximum torque production capacity in flexion is at 0° (Otis et al., Reference Otis, Warren, Backus, Santner and Mabrey1990; Koski and McGill, Reference Koski and McGill1994). Thus, ULE supports the shoulder where it is most needed. However, if the elbow is flexed, the shoulder torque is reduced and the profile changes. For an elbow flexed at 90°, the maximum shoulder torque was obtained at 60° (Kim et al., Reference Kim, Kim, Jung, Lee and Bae2022). Taking the example of a crate lifting task, the torque–angle relationship of exoskeletons was most effective for peak torques occurring between 75 and 105° (Seiferheld et al., Reference Seiferheld, Frost, Krog, Skals and Andersen2022).

However, adjusting the level of assistance of ULEs remains challenging because the shoulder torques vary depending on the task. For overhead drilling tasks with a drill weighting about 2 kg, average shoulder torques of 10 Nm with a 90° elbow, and 30 Nm with an extended elbow were calculated (Anton et al., Reference Anton, Shibley, Fethke, Hess, Cook and Rosecrance2001). For another drilling task with a 2-kg tool, shoulder torques between 6 and 11 Nm were estimated (Panariello et al., Reference Panariello, Grazioso, Caporaso, Palomba, Di Gironimo and Lanzotti2022). For two-handed lifting tasks with weights of 5, 10, and 15 kg, the 95th percentiles of shoulder flexion torques were 18.7, 31.7 and 44.3 Nm respectively, and 19.7 Nm for one-handed OHW lifting tasks with a weight of 3 kg (Huysamen et al., Reference Huysamen, Power and O’Sullivan2020). The setting of the assistance could therefore be specific to the type of task (movement, weight handled, etc.). This specific setting may not be suitable for a sequence of tasks carried out in an occupational environment. A balance must be found between insufficient and excessive assistance (Sylla et al., Reference Sylla, Bonnet, Colledani and Fraisse2014; Van Engelhoven et al., Reference Van Engelhoven, Poon, Kazerooni, Barr, Rempel and Harris-Adamson2018; Sänger et al., Reference Sänger, Yao, Schubert, Wolf, Molz, Miehling, Wartzack, Gwosch, Matthiesen and Weidner2022). A recent study recommended two different settings for static and dynamic pointing tasks. A maximum assistance (i.e., 6 Nm) was recommended for static tasks and a moderate level (60 or 80% of the maximum) for dynamic tasks so as not to increase the activation of antagonist muscles when lowering the arm (Ramella et al., Reference Ramella, Grazi, Giovacchini, Trigili, Vitiello and Crea2024). The operation of ULE can be counterproductive for shoulder extension movements due to the opposing torque and hysteresis of these devices. Consequently, these devices should be used within their angular ranges of assistance for tasks involving holding the arms at shoulder height or overhead lifting. It is also essential to consider the length of time for which the ULE provides appropriate assistance. Furthermore, a potential solution to address the issues related to adjusting the level of assistance could involve real-time modulation through active or semiactive systems (Grazi et al., Reference Grazi, Trigili, Proface, Giovacchini, Crea and Vitiello2020; Crea et al., Reference Crea, Beckerle, Looze, Pauw, Grazi, Kermavnar, Masood, O’Sullivan, Pacifico, Rodriguez-Guerrero, Vitiello, Ristić-Durrant and Veneman2021).

5. Conclusion

The characterization of the ULE torques under static and dynamic conditions can be used to adapt the settings of the exoskeleton according to the individual characteristics of the user and the mechanical characteristics of the tasks. Going further toward individualized settings, the mechanical properties of the exoskeleton’s assistive torques, such as hysteresis, were modeled using equations with the exoskeleton’s assistance level and opening angle as inputs. In this way, the optimal setting can be assessed by simulation, striking a balance between the benefits and undesirable effects of wearing ULE.

Supplementary material

The supplementary material for this article can be found at http://doi.org/10.1017/wtc.2025.8.

Data availability statement

The datasets generated during and/or analyzed during this study are available from the corresponding author on reasonable request.

Acknowledgments

The authors thank Olivier Morel for setting up the test bench.

Author contribution

K.D., C.H., E.R., C.P., and L.C. conceived the study. K.D. and E.R. designed the study and conducted data gathering. E.R. and I.C.-U. processed data. All authors analyzed the data. E.R. wrote the article. K.D., C.H., C.P., I.C.-U., and L.C. reviewed and edited the article.

Funding statement

This research received no specific grant from any funding agency, commercial or not-for-profit sectors.

Competing interests

The authors declare none.

References

Akaike, H (1974) A new look at the statistical model identification. IEEE Transactions on Automatic Control 19(6), 716723. https://doi.org/10.1109/TAC.1974.1100705.CrossRefGoogle Scholar
Anton, D, Shibley, LD, Fethke, NB, Hess, J, Cook, TM and Rosecrance, J (2001) The effect of overhead drilling position on shoulder moment and electromyography. Ergonomics 44(5), 489501. https://doi.org/10.1080/00140130120079.CrossRefGoogle ScholarPubMed
Asgari, M, Phillips, EA, Dalton, BM, Rudl, JL and Crouch, DL (2020, June 2) Design and performance evaluation of a wearable passive cable-driven shoulder exoskeleton bioRxiv 96453. https://doi.org/10.1101/2020.05.29.096453.CrossRefGoogle Scholar
Barthelme, J, Sauter, M, Mueller, C and Liebers, F (2021) Association between working in awkward postures, in particular overhead work, and pain in the shoulder region in the context of the 2018 BIBB/BAuA employment survey. BMC Musculoskeletal Disorders 22(1), 624. https://doi.org/10.1186/s12891-021-04482-4.CrossRefGoogle ScholarPubMed
Cohen, J (1992) A power primer. Psychological Bulletin 112(1), 155159. https://doi.org/10.1037//0033-2909.112.1.155.CrossRefGoogle ScholarPubMed
Crea, S, Beckerle, P, Looze, MD, Pauw, KD, Grazi, L, Kermavnar, T, Masood, J, O’Sullivan, LW, Pacifico, I, Rodriguez-Guerrero, C, Vitiello, N, Ristić-Durrant, D and Veneman, J (2021) Occupational exoskeletons: A roadmap toward large-scale adoption. Methodology and challenges of bringing exoskeletons to workplaces. Wearable Technologies 2, e11. https://doi.org/10.1017/wtc.2021.11.CrossRefGoogle ScholarPubMed
de Araujo Ribeiro Alvares, JB, Rodrigues, R, de Azevedo Franke, R, da Silva, BGC, Pinto, RS, Vaz, MA and Baroni, BM (2015) Inter-machine reliability of the biodex and cybex isokinetic dynamometers for knee flexor/extensor isometric, concentric and eccentric tests. Physical Therapy in Sport 16(1), 5965. https://doi.org/10.1016/j.ptsp.2014.04.004.CrossRefGoogle ScholarPubMed
de Leva, P (1996) Adjustments to zatsiorsky-seluyanov’s segment inertia parameters. Journal of Biomechanics 29(9), 12231230. https://doi.org/10.1016/0021-9290(95)00178-6.CrossRefGoogle ScholarPubMed
de Vries, A, Murphy, M, Könemann, R, Kingma, I and de Looze, M (2019) The amount of support provided by a passive arm support exoskeleton in a range of elevated arm postures. IISE Transactions on Occupational Ergonomics and Human Factors 7(3), 311321. https://doi.org/10.1080/24725838.2019.1669736.CrossRefGoogle Scholar
Delgado-Llamas, A, Marín-Boné, J and Marín-Zurdo, JJ (2023) Can we simulate the biomechanical effects of exoskeletons prior to workstation implementation? Application of the forces ergonomic method. International Journal of Industrial Ergonomics 94, 103409. https://doi.org/10.1016/j.ergon.2023.103409.CrossRefGoogle Scholar
Estrázulas, JA, Estrázulas, JA, de Jesus, K, de Jesus, K, da Silva, RA and Libardoni dos Santos, JO (2020) Evaluation isometric and isokinetic of trunk flexor and extensor muscles with isokinetic dynamometer: A systematic review. Physical Therapy in Sport 45, 93102. https://doi.org/10.1016/j.ptsp.2020.06.008.CrossRefGoogle ScholarPubMed
Exoskeleton Report T (2022, April 22) Ottobock shoulder [Exoskeleton Report]. Available at https://exoskeletonreport.com/product/ottobock-shoulder/.Google Scholar
Finke, LB, Kroll-Orywahl, O, Tüttemann, M, Bremer, M, Merker, MA, Mizera, O, Mladek, M, Röschel, F, Schirrmeister, B, Vogel, C, Bornmann, J, Lanzinger, M, Käsmann, S and Schwedhelm, P (2022, Device for supporting at least one arm of a user and for supporting at least one upper arm (U.S. Pat. No. 20220031551A2). Available at https://patents.google.com/patent/US20220031551A2/en.Google Scholar
Grazi, L, Trigili, E, Proface, G, Giovacchini, F, Crea, S and Vitiello, N (2020) Design and experimental evaluation of a semi-passive upper-limb exoskeleton for workers with motorized tuning of assistance. IEEE Transactions on Neural Systems and Rehabilitation Engineering 28(10), 22762285. https://doi.org/10.1109/TNSRE.2020.3014408.CrossRefGoogle ScholarPubMed
Habets, B, Staal, JB, Tijssen, M and van Cingel, R (2018) Intrarater reliability of the humac NORM isokinetic dynamometer for strength measurements of the knee and shoulder muscles. BMC Research Notes 11(1), 15. https://doi.org/10.1186/s13104-018-3128-9.CrossRefGoogle ScholarPubMed
Harbo, T, Brincks, J and Andersen, H (2012) Maximal isokinetic and isometric muscle strength of major muscle groups related to age, body mass, height, and sex in 178 healthy subjects. European Journal of Applied Physiology 112(1), 267275. https://doi.org/10.1007/s00421-011-1975-3.CrossRefGoogle ScholarPubMed
Hartmann, VN, Rinaldi, DDM, Taira, C and Forner-Cordero, A (2021) Industrial upper-limb exoskeleton characterization: Paving the way to new standards for benchmarking. Machines 9(12), 362 https://doi.org/10.3390/machines9120362.CrossRefGoogle Scholar
Hurvich, CM, Simonoff, JS and Tsai, C-L (1998) Smoothing parameter selection in nonparametric regression using an improved akaike information criterion. Journal of the Royal Statistical Society Series B: Statistical Methodology 60(2), 271293. https://doi.org/10.1111/1467-9868.00125.CrossRefGoogle Scholar
Huysamen, K, Power, V and O’Sullivan, L (2020) Kinematic and kinetic functional requirements for industrial exoskeletons for lifting tasks and overhead lifting. Ergonomics 63, 7 818830. https://doi.org/10.1080/00140139.2020.1759698.CrossRefGoogle ScholarPubMed
Martin Bland, J & Altman, D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet 327(8476), 307310. https://doi.org/10.1016/S0140-6736(86)90837-8.CrossRefGoogle Scholar
Kim, J, Kim, J, Jung, Y, Lee, D and Bae, J (2022) A passive upper limb exoskeleton with tilted and offset shoulder joints for assisting overhead tasks IEEE/ASME Transactions on Mechatronics 27(6), 49634973. https://doi.org/10.1109/TMECH.2022.3169617.CrossRefGoogle Scholar
Koopman, AS, Kingma, I, Faber, GS, de Looze, MP and van Dieën, JH (2019) Effects of a passive exoskeleton on the mechanical loading of the low back in static holding tasks. Journal of Biomechanics 83, 97103. https://doi.org/10.1016/j.jbiomech.2018.11.033.CrossRefGoogle Scholar
Koski, AV and McGill, SM (1994) Dynamic shoulder flexion strength: For use in occupational risk analysis and clinical assessment. Clinical Biomechanics 9(2), 99104. https://doi.org/10.1016/0268-0033(94)90031-0.CrossRefGoogle ScholarPubMed
Kranenborg, SE, Greve, C, Reneman, MF and Roossien, CC (2023) Side-effects and adverse events of a shoulder- and back-support exoskeleton in workers: A systematic review. Applied Ergonomics 111, 104042. https://doi.org/10.1016/j.apergo.2023.104042.CrossRefGoogle ScholarPubMed
Linnenberg, C and Weidner, R (2022) Industrial exoskeletons for overhead work: Circumferential pressures on the upper arm caused by the physical human-machine-interface. Applied Ergonomics 101, 103706. https://doi.org/10.1016/j.apergo.2022.103706.CrossRefGoogle ScholarPubMed
Looft, JM and Law, LAF (2015) Modelling three-dimensional human strength capacity: Logistic vs. polynomial surface equations. International Journal of Human Factors Modelling and Simulation 5(1), 5 18. https://doi.org/10.1504/IJHFMS.2015.068122.CrossRefGoogle Scholar
Ma, T, Zhang, Y, Choi, SD and Xiong, S (2023) Modelling for design and evaluation of industrial exoskeletons: A systematic review. Applied Ergonomics 113, 104100. https://doi.org/10.1016/j.apergo.2023.104100.CrossRefGoogle ScholarPubMed
Madinei, S, Kim, S, Park, J-H, Srinivasan, D and Nussbaum, MA (2022) A novel approach to quantify the assistive torque profiles generated by passive back-support exoskeletons. Journal of Biomechanics 145, 111363. https://doi.org/10.1016/j.jbiomech.2022.111363.CrossRefGoogle ScholarPubMed
McCann, CM, Hohimer, CJ, O’Neill, CT, Young, HT, Bertoldi, K and Walsh, CJ (2023) In-situ measurement of multi-axis torques applied by wearable soft robots for shoulder assistance IEEE Transactions on Medical Robotics and Bionics 5(2), 363374. https://doi.org/10.1109/TMRB.2023.3258499.CrossRefGoogle Scholar
Moeller, T, Krell-Roesch, J, Woll, A and Stein, T (2022) Effects of upper-limb exoskeletons designed for use in the working environment—a literature review. Frontiers in Robotics and AI 9, 858893 https://doi.org/10.3389/frobt.2022.858893.CrossRefGoogle ScholarPubMed
Ostertagová, E (2012) Modelling using polynomial regression. Procedia Engineering 48, 500506. https://doi.org/10.1016/j.proeng.2012.09.545.CrossRefGoogle Scholar
Otis, JC, Warren, RF, Backus, SI, Santner, TJ and Mabrey, JD (1990) Torque production in the shoulder of the normal young adult male: The interaction of function, dominance, joint angle, and angular velocity The American Journal of Sports Medicine 18(2), 119123. https://doi.org/10.1177/036354659001800201.CrossRefGoogle ScholarPubMed
Panariello, D, Grazioso, S, Caporaso, T, Palomba, A, Di Gironimo, G and Lanzotti, A (2022) Biomechanical analysis of the upper body during overhead industrial tasks using electromyography and motion capture integrated with digital human models. International Journal on Interactive Design and Manufacturing (IJIDeM), 16(2), 733752. https://doi.org/10.1007/s12008-022-00862-9.CrossRefGoogle Scholar
Ramella, G, Grazi, L, Giovacchini, F, Trigili, E, Vitiello, N and Crea, S (2024) Evaluation of antigravitational support levels provided by a passive upper-limb occupational exoskeleton in repetitive arm movements. Applied Ergonomics 117, 104226. https://doi.org/10.1016/j.apergo.2024.104226.CrossRefGoogle ScholarPubMed
Rossini, M, De Bock, S, van der Have, A, Flynn, L, Rodriguez-Cianca, D, De Pauw, K, Lefeber, D, Geeroms, J and Rodriguez-Guerrero, C (2021) Design and evaluation of a passive cable-driven occupational shoulder exoskeleton IEEE Transactions on Medical Robotics and Bionics 3(4), 10201031. https://doi.org/10.1109/TMRB.2021.3110679.CrossRefGoogle Scholar
Sänger, J, Yao, Z, Schubert, T,Wolf, A, Molz, C, Miehling, J,Wartzack, S, Gwosch, T, Matthiesen, S and Weidner, R (2022) Evaluation of active shoulder exoskeleton support to deduce application-oriented optimization potentials for overhead work. Applied Sciences 12(21), 10805. https://doi.org/10.3390/app122110805.CrossRefGoogle Scholar
Scherb, D,Wartzack, S and Miehling, J (2023) Modelling the interaction between wearable assistive devices and digital human models—a systematic review. Frontiers in Bioengineering and Biotechnology 10, 1044275. https://doi.org/10.3389/fbioe.2022.1044275.CrossRefGoogle ScholarPubMed
Seiferheld, BE, Frost, J, Krog, M, Skals, S and Andersen, MS (2022) Biomechanical investigation of a passive upper-extremity exoskeleton for manual material handling – a computational parameter study and modelling approach. International Journal of Human Factors Modelling and Simulation 7(3), 275300. https://doi.org/10.1504/IJHFMS.2022.124304.CrossRefGoogle Scholar
Sugiura, N (1978) Further analysis of the data by akaike’s information criterion and the finite corrections: Further analysis of the data by akaike’s Communications in Statistics – Theory and Methods 7(1), 1326. https://doi.org/10.1080/03610927808827599.CrossRefGoogle Scholar
Sylla, N, Bonnet, V, Colledani, F and Fraisse, P (2014) Ergonomic contribution of ABLE exoskeleton in automotive industry. International Journal of Industrial Ergonomics 44(4), 475481. https://doi.org/10.1016/j.ergon.2014.03.008.CrossRefGoogle Scholar
Theurel, J and Desbrosses, K (2019) Occupational exoskeletons: Overview of their benefits and limitations in preventing work-related musculoskeletal disorders. IISE Transactions on Occupational Ergonomics and Human Factors 7(3), 264280. https://doi.org/10.1080/24725838.2019.1638331.CrossRefGoogle Scholar
van der Have, A, Rossini, M, Rodriguez-Guerrero, C, Van Rossom, S and Jonkers, I (2022) The exo4work shoulder exoskeleton effectively reduces muscle and joint loading during simulated occupational tasks above shoulder height. Applied Ergonomics 103, 103800. https://doi.org/10.1016/j.apergo.2022.103800.CrossRefGoogle ScholarPubMed
Van Engelhoven, L, Poon, N, Kazerooni, H, Barr, A, Rempel, D and Harris-Adamson, C (2018) Evaluation of an adjustable support shoulder exoskeleton on static and dynamic overhead tasks. Proceedings of the Human Factors and Ergonomics Society Annual Meeting 62(1), 804808. https://doi.org/10.1177/1541931218621184.CrossRefGoogle Scholar
Van Engelhoven, L, Poon, N, Kazerooni, H, Rempel, D, Barr, A and Harris-Adamson, C (2019) Experimental evaluation of a shoulder-support exoskeleton for overhead work: Influences of peak torque amplitude, task, and tool mass. IISE Transactions on Occupational Ergonomics and Human Factors 7(3), 250263. https://doi.org/10.1080/24725838.2019.1637799.CrossRefGoogle Scholar
Van Harmelen, V and Schnieders, J (2022) Measuring the amount of support of lower back exoskeletons. Available at https://www.laevo-exoskeletons.com/en/learning-centerGoogle Scholar
Wagenmakers, E-J and Farrell, S (2004) AIC model selection using akaike weights. Psychonomic Bulletin & Review 11(1), 192196. https://doi.org/10.3758/BF03206482.CrossRefGoogle ScholarPubMed
Watterworth, MWB, Dharmaputra, R, Porto, R, Cort, JA and La Delfa, NJ (2023) Equations for estimating the static supportive torque provided by upper-limb exoskeletons. Applied Ergonomics 113, 104092. https://doi.org/10.1016/j.apergo.2023.104092.CrossRefGoogle ScholarPubMed
Whitcomb, LJ, Kelley, MJ, & Leiper, CI. (1995). A comparison of torque production during dynamic strength testing of shoulder abduction in the coronal plane and the plane of the scapula. The Journal of Orthopaedic and Sports Physical Therapy, 21(4), 227232. https://doi.org/10.2519/jospt.1995.21.4.227.CrossRefGoogle ScholarPubMed
Winter, D (2009) Biomechanics and Motor Control of Human Movement, 4th Edn. John Wiley & Sons, Ltd, pp. 4581. https://doi.org/10.1002/9780470549148.ch3.CrossRefGoogle Scholar
Zhou, X and Zheng, L (2021) Model-based comparison of passive and active assistance designs in an occupational upper limb exoskeleton for overhead lifting IISE Transactions on Occupational Ergonomics and Human Factors 9(3), 167185. https://doi.org/10.1080/24725838.2021.1954565.CrossRefGoogle Scholar
Figure 0

Figure 1. Ottobock shoulder exoskeleton with its different elements and assistance levels to the study: minimum in light blue (Min), two intermediate levels in green and magenta (Int 1 and Int 2), and a maximum in dark blue (Max) (Exoskeleton Report, 2022).

Figure 1

Figure 2. Detailed view of the test bench with the Ottobock shoulder exoskeleton positioned in its maximum opening angle (155°) on the HUMAC NORM isokinetic dynamometer.

Figure 2

Figure 3. Torque–angle relationship of the Ottobock shoulder under different conditions: STA (left), FLE (center), and EXT (right) for the four assistance levels (Min, Int1, Int2, Max), represented by the mean with the standard deviation in the band. The dynamic conditions FLE and EXT are represented for the 12 angular velocities from 20 to 240°/s.

Figure 3

Figure 4. Boxplot representation to compare the exoskeleton’s maximum torques (top) and their angular positions (bottom) for the four assistance levels (Min, Int1, Int2, and Max) in STA, FLE, and EXT including 5 trials and 12 angular velocities.

Figure 4

Table 1. Ottobock shoulder assistive torques equations fitted by polynomial regression models for STA, FLE, and EXT conditions. The equations were z = f(x,y), where x corresponds to the considered angle and y the assistance levels between 1 and 4, where 1 is the Min level, 2 and 3 Int1 and Int2 levels, and 4 the Max level, with their fitting indicators, RMSE and $ {\mathrm{R}}_{\mathrm{adj}}^2 $

Supplementary material: File

Ricard et al. supplementary material 1

Ricard et al. supplementary material
Download Ricard et al. supplementary material 1(File)
File 7.9 MB
Supplementary material: File

Ricard et al. supplementary material 2

Ricard et al. supplementary material
Download Ricard et al. supplementary material 2(File)
File 557.4 KB