Published online by Cambridge University Press: 21 February 2011
A technique for observing both plan view and cross sections of a specimen directly in a transmission electron microscope (TEM) without relying on a tedious thinning operation was developed. This technique involves cleaving a specimen perpendicular to the plane, so that the thin (electron transparent) section of the cleaved edge can be directly imaged by TEM. The only limitations of this technique are that a specimen must be readily criacked or cleaved and that, since the transparent region is often bounded by a 90° corner, the extent of electron transparent region is somewhat localized. Nevertheless, the technique has the advantages of the ease of specimen preparation, and the absence of contamination or damage introduced in other conventional thinning methods. The geometry of the cleaved specimen is also suitable for reflection electron microscopy.